PUMP
REFILL:
- Drain
residual volume completely (verify actual amount drained to amount
calculated by computer for accuracy, and to monitor system)
- check
for drug leak and air leak in tubing connection setup

- enter
reservoir port, not side access port.
- if
unabe to locate or enter reservoir port use template for location.
- if
unsure - do not refill - placing medication into access can result
in medication being injected directly into catheter - and can
cause overdose - this will inject directly into subarachnoid
space and bypass pump reservoir. This is helpful for dye diagnostics
to check integrity of line. This can not be done in usual refill
situation due to the different size needle (smaller) that is
required to bypass mesh over access port for entry - the larger
needle in refill kit for entry into reservoir port will not be
allowed to pass mesh into the access port even in error.
- at
the present time refills are full serve, there is no self-serve
option.
- if
refill is considered to be painful by patient, use emla cream
applied over pump 1 hour prior to refill. this provides topical
local anesthetic. this is wiped off and cleaned prior to refill.
the cream makes the refill procedure painless.
- DO
NOT OVERFILL

- ~PRESSURE
CHECK WITH MANOMETER (must place manometer level with or below
pump for optimum pressure check)

Medication
refills are given through reservoir port, pump diagnostics (if
problem occurs) are done via access port (check pump line for
blockage, assess patency, "pump shunt-o-gram") under
flouroscopy. Mesh over access port requires smaller gauge needle
to enter than reservoir port. This size differential prevents
inadvertent entry of larger needle used during refill into reservoir
port from incorrectly being placed and entering access port.

