SynchroMed
EL
INFUSION
SYSTEM
SynchroMed
Pump Enhancements
Special Features
- Extended battery
Life
- from 4 - 7 years
when programmed to an average flow rate of 500 ul/day
- Lower flow rates
- from 96 ul down
to 48 ul per day (+/- 15%) allowing more therapeutic options
- Non-glare matte
finish
- Extra suture
loop on models with loops
- No changes in
clinical implant procedure or pump programming
Specifications
for SynchroMed EL Infusion System
Pump Models
|
|
10 cc |
18 cc |
10 cc w/cap |
18 cc w/cap |
|
Mass (empty
max) |
165 g (5.8
oz) |
185 g (6.5
oz) |
185 g (6.5
oz) |
205 g (7.2 oz) |
|
Diameter |
70.4mm (max)
(2.77in) |
70.4mm (max)
(2.77in) |
85.2mm (nom)
(3.36in) |
85.2mm (nom) (3.36in) |
|
Thickness |
22.86mm (.90in) |
27.5mm (1.08in) |
22.86mm (.90in) |
27.5mm (1.08in) |
|
Reservoir
Capacity |
10 mL |
18 mL |
10 mL |
18 mL |
- Refill Port
(septum)
- Material: Silicone
rubber
- Puncture Life:
500 punctures
- (22-gauge Huber-type
needle)
- Fluid Output
- Maximum Infusion
Rate: 900 ul/hr; 0.9mL/hr
- Minimum Infusion
Rate: 48 ul/day; 0.048mL/day
- Bacterial Retentive
Filter
- Power Source
- Battery: Lithium
thionyl-chloride
- Life: 6 - 7
years (rate dependent)
- Typical: 7 years
at 0.5 mL/day
- Screened Catheter
Access Port
- Catheter can
be accessed directly with a 25-gauge needle for drug administration,
cerebrospinal fluid (CSF) sampling, diagnosis, or troubleshooting.
- Screen allows
25-gauge (or smaller) needles to pass through the septum.
- Screen prevents
inadvertent catheter access with a 22-gauge (or larger) needle
during pump refill procedure.
- Funnel-shaped
holes in the screen direct the 25-gauge needle into the catheter
access port septum.